Insulation Resistance up to 10TΩ | |
• | Short-Circuit Current up to 5mA |
• | Wide Test Voltage from 250V to 5000V |
• | Diagnostic Insulation Tests: IR, PI, DAR, DD, SV, RAMP (Breakdown and Burn modes), Capacitance, Leakage Current |
• | Wireless communication by Bluetooth for transferring and showing real-time data to PC and Android device |
• | Memory and Logging functions |
• | Filter function reduces noise interference for obtaining stable measurements |
• | Large back light display with Bar-Graph indication |
• | Robust design for field use with IP65 (lid closed) |
• | Powered by rechargeable battery |
• | Safety Standard IEC 61010-1 CAT.Ⅳ 600V |
Insulation resistance | |||||||||||||||||||||||||||||||
Test voltage | 250V (*1) | 500V | 1000V | 2500V | 5000V | ||||||||||||||||||||||||||
Max measurement value |
9.99GΩ | 99.9GΩ | 199GΩ | 999GΩ | 9.99TΩ | ||||||||||||||||||||||||||
Accuracy | 0.0 - 99.9MΩ ±5%rdg±3dgt |
0.0 - 999MΩ ±5%rdg±3dgt |
0.0 - 1.99GΩ ±5%rdg±3dgt |
0.0 - 99.9GΩ ±5%rdg±3dgt |
0.0 - 99.9GΩ ±5%rdg±3dgt |
||||||||||||||||||||||||||
0.1G - 9.99GΩ ±20%rdg |
1G - 99.9GΩ ±20%rdg |
2G - 199GΩ ±20%rdg |
100G - 999GΩ ±20%rdg |
0.1T - 9.99TΩ ±20%rdg |
|||||||||||||||||||||||||||
Short circuit current | Max 5.0mA | ||||||||||||||||||||||||||||||
Output voltage | Accuracy | -10 - +10% | -10 - +20% | 0 - +20% | |||||||||||||||||||||||||||
Variable | - | -20% - 0% (5% step) | |||||||||||||||||||||||||||||
Monitor | ±10%rdg±20V | ||||||||||||||||||||||||||||||
Voltage measurement | |||||||||||||||||||||||||||||||
Measuring range | AC: 30 - 600V (50/60Hz), DC: ±30 - ±600V | ||||||||||||||||||||||||||||||
Accuracy | ±2%rdg±3dgt | ||||||||||||||||||||||||||||||
Current measurement | |||||||||||||||||||||||||||||||
Measuring range | 0.00nA - 5.50mA | ||||||||||||||||||||||||||||||
Accuracy | ±10%rdg (*2) | ||||||||||||||||||||||||||||||
Capacitance measurement | |||||||||||||||||||||||||||||||
Measuring range | 5.0nF - 50.0µF | 5.0nF - 25.0µF | |||||||||||||||||||||||||||||
Accuracy | ±5%rdg±5dgt | ||||||||||||||||||||||||||||||
General | |||||||||||||||||||||||||||||||
Power source | Rechargeable Battery (Lead-acid Battery) 12V (*3) Charging power: DC 15VA MAX |
||||||||||||||||||||||||||||||
Communication Interface | Bluetooth: Ver2.1 + EDR Class2, USB: Ver1.1 | ||||||||||||||||||||||||||||||
Applicable standards | IEC 61010-1, 61010-2-030 CAT.Ⅳ 600V Pollution degree 2, IEC 61010-031, IEC 61326-1, 2-2 |
||||||||||||||||||||||||||||||
Dimension | 208 (L) × 225 (W) × 130 (D) mm 380 (L) × 430 (W) × 154 (D) mm (Hard case) |
||||||||||||||||||||||||||||||
Weight | 3127: 4kg Approx. (including battery), Total: 8kg Approx. (including Accessories) |
||||||||||||||||||||||||||||||
Accessories | 7165A (Line probe), 7224A (Earth cord), 7225A (Guard cord), 8019 (Hook type prod), 8254 (Straight type prod), 8327EU (Power adaptor 15V/1A), 9171 (Hard case), Instruction manual, Calibration Certificate |
||||||||||||||||||||||||||||||
Optional | 7168A (Line probe with alligator clip: 3m), 7253 (Longer line probe with alligator clip: 15m), 8258 (USB communication set), 8302 (Adaptor for recorder 1mV/1µA) |
||||||||||||||||||||||||||||||
(*1): | IR mode only |
(*2): | Determined by resistance and Voltage values (over 10MΩ) |
(*3): | No measurements are possible while charging |